Certification/ Class (Standard)
|
N95 (NIOSH-42C
FR84)
|
FFP2
(EN 149-2001)
|
KN95
(GB2626-20 06)
|
P2
(AS/NZ 1716:2012)
|
Korea 1st Class (KMOEL - 2017-64)
|
DS (Japan JMHLW-
Notification 214, 2018)
|
Filter performance – (must be ≥ X% efficient)
|
≥
95%
|
≥
94%
|
≥
95%
|
≥
94%
|
≥
94%
|
≥
95%
|
Test agent
|
NaCl
|
NaCl
and paraffin oil
|
NaCl
|
NaCl
|
NaCl
and paraffin oil
|
NaCl
|
Flow rate
|
85
L/min
|
95
L/min
|
85
L/min
|
95
L/min
|
95
L/min
|
85
L/min
|
Total inward leakage (TIL)* – tested on human subjects each
performing exercises
|
N/A
|
≤
8% leakage (arithmetic mean)
|
≤
8% leakage (arithmetic mean)
|
≤
8% leakage (individual and arithmetic mean)
|
≤
8% leakage (arithmetic mean)
|
Inward
Leakage measured and included in User Instructions
|
Inhalation resistance – max pressure drop
|
≤
343 Pa
|
≤ 70 Pa (at 30 L/min)
≤ 240 Pa (at 95 L/min)
≤ 500 Pa (clogging)
|
≤
350 Pa
|
≤ 70 Pa (at 30 L/min)
≤ 240 Pa (at 95 L/min)
|
≤ 70 Pa (at 30 L/min)
≤ 240 Pa (at 95 L/min)
|
≤ 70 Pa (w/valve)
≤ 50 Pa (no valve)
|
Flow rate
|
85
L/min
|
Varied
– see above
|
85
L/min
|
Varied
– see above
|
Varied
– see above
|
40
L/min
|
Exhalation resistance - max pressure drop
|
≤
245 Pa
|
≤
300 Pa
|
≤
250 Pa
|
≤
120 Pa
|
≤
300 Pa
|
≤ 70 Pa (w/valve)
≤ 50 Pa (no valve)
|
Flow rate
|
85
L/min
|
160
L/min
|
85
L/min
|
85
L/min
|
160
L/min
|
40
L/min
|
Exhalation valve leakage requirement
|
Leak
rate ≤ 30 mL/min
|
N/A
|
Depressurizatio
n to 0 Pa ≥ 20 sec
|
Leak
rate ≤ 30 mL/min
|
visual
inspection after 300 L /min for 30 sec
|
Depressurizatio
n to 0 Pa ≥ 15 sec
|
Force applied
|
-245
Pa
|
N/A
|
-1180
Pa
|
-250
Pa
|
N/A
|
-1,470
Pa
|
CO2 clearance requirement
|
N/A
|
≤
1%
|
≤
1%
|
≤
1%
|
≤
1%
|
≤
1%
|
*Japan JMHLW-Notification 214 requires an Inward Leakage test
rather than a TIL test.
|
|
|
|
|
|
|
|